Service · OQ filter test units OQ for filter test units in autoclaves and freeze dryers
We check how reliably your system tests its filters.
As part of operational qualification (OQ), we assess your system’s integrated filter test unit. We verify whether it measures water flow during the water intrusion test with sufficient accuracy and reproducibility, and whether it correctly detects simulated fault conditions. Testing is performed under on-site conditions, with clear documentation of the procedures and results.
At a glance
- Qualification of integrated filter test units
- For autoclaves and freeze dryers
- Single and double filter systems with pre-filter and main filter
- Comparative measurements with calibrated reference instruments
- Test plan, results and evaluation in a single documentation package
What we test
The specific scope of testing is tailored to the filter test unit of your system.
Pressure sensors
Review of the available calibration certificates and the documented measurement deviations against the defined acceptance criteria.
Leak tightness
Leak test of the filter systems or of the filter simulators used, in accordance with the test plan.
Hardware volume
Determination of the volume relevant to the measurement using a method suited to the system design. Additional volume in pipework and connections is taken into account.
Water flow measurement
Verification with filter simulator, drip nozzle and analytical balance. Repeated measurements show the deviations from the independent reference.
Test with filter cartridges
Water intrusion test with the filter cartridges intended for operation – depending on the system on the single filter or on pre-filter and main filter.
Fault detection
Simulation of defined fault conditions, for example a filter breakthrough. We check whether the system detects the fault and triggers the intended error message or abort sequence.
Test pressure control
Comparison of the actual test pressure with the specified set point over several test runs.
Analysis and evaluation
Evaluation of the results against defined acceptance criteria, documentation of any deviations and overall assessment of the filter test unit.
How the integrated filter test unit works
In the water intrusion test the hydrophobic vent filter is tested in place. The plant control system calculates the water flow from the measured pressure drop and the underlying gas volume. For meaningful results, volume, pressure and temperature, among other things, must be taken into account.
- Heatable filter housing with vent filter
- Pressure sensor and temperature sensor on top of the filter housing
- Condensate treatment vessel (test water)
- Precision pressure regulator with gauge
- Pneumatic valves (filling, drain and inlet valve)
- Cooling water inlet and outlet
Example: test sequence on a freeze dryer
- Prepare test waterClean steam condenses in the cooled treatment vessel and provides the test water.
- Set the temperatureThe test water is cooled to the temperature specified for the test.
- Fill the filter housingThe water rises from below in the housing. As soon as the filter membrane is completely covered, the remaining gas volume above the membrane is trapped.
- Stabilise the test pressureThe precision pressure regulator sets the specified test pressure. The defined stabilisation phase runs before the measurement.
- Determine the water flowAfter the filling valve has closed, the system records the pressure drop and calculates the water flow from it.
- Drain and pre-dryThe test water is drained. Reverse venting through the filter supports pre-drying.
The specific set-up and sequence depend on the individual system.
For single and double filter systems
We qualify filter test units with a single filter as well as systems with pre-filter and main filter. We take the respective system design into account and determine the hardware volume relevant to the measurement.
For verification we use filter simulators developed in-house. A drip nozzle allows water to be withdrawn in a controlled way during the measuring phase. We determine the amount of water withdrawn with an analytical balance and compare it with the result of the system.
In this way we check the measurement against an independent reference – also in the range of the filter test limit.

Documented for traceability
You receive a consolidated OQ documentation package containing the test plan, measured values, evaluation and supporting records.
Test plan
Test objectives, scope, prerequisites and requirements for the measuring equipment used.
Method and principles
Description of the filter test method, the reference measurements and the relevant physical principles.
Test results and evaluation
Measured values, acceptance criteria, documented deviations and signatures of the persons performing and reviewing the tests.
Records and attachments
List of measuring equipment, calibration documents, test reports and relevant documents on the filters used.